FDA Device recall Z-2019-2012
Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or…
- FDA Device
- Class II
- terminated
- Published 2012-07-25
On 2012-07-25 the FDA classified a Class II recall of Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or… by Spacelabs Healthcare, citing there were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will dr….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2019-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-06-12 |
| Published | 2012-07-25 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Spacelabs Healthcare |
| Distribution | FL, IA, OK |
Product
Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other devices via Flexport interfaces. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Spacelabs Healthcare Xprezzon Bedside Monitor may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitor allows network based applications to open windows and display information the Xprezzon and other networked monitors. The Spacelabs Healthcare Xprezzon Bedside Monitor is designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Xprezzon Bedside Monitors is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.
Reason
There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.
Identifiers
| code_info | 040-1632-00; 1393-000013; 1393-000039; 1393-000040; 1393-000041; 1393-000341; 1393-000382; 1393-000383; 1393-000384; 13…040-1632-00; 1393-000013; 1393-000039; 1393-000040; 1393-000041; 1393-000341; 1393-000382; 1393-000383; 1393-000384; 1393-000385; 1393-000386; 1393-000387; 1393-000388; 1393-000808; 1393-000831; 1393-001374; 1393-001375; 1393-001453; 1393-001990; 1393-001991; and 1393-001992. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Spacelabs Healthcare
- Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelab…
FDA Device · Class II · 2026-08-19 · Spacelabs Healthcare - Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A Product…
FDA Device · Class II · 2026-05-13 · Spacelabs Healthcare - Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
FDA Device · Class II · 2026-04-15 · Spacelabs Healthcare - Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology In…
FDA Device · Class II · 2025-11-12 · Spacelabs Healthcare - Brand Name: Branding is not applied to this product
FDA Device · Class II · 2025-10-22 · Spacelabs Healthcare
All Spacelabs Healthcare recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2019-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.