FDA Device recall Z-0167-2026
Brand Name: Branding is not applied to this product
- FDA Device
- Class II
- ongoing
- Published 2025-10-22
On 2025-10-22 the FDA classified a Class II recall of Brand Name: Branding is not applied to this product by Spacelabs Healthcare, citing due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0167-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-09-05 |
| Published | 2025-10-22 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Spacelabs Healthcare |
| Distribution | US |
Product
Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393
Reason
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
Identifiers
| code_info | Lot Code: UDI-DI for products that are used with the display cables. (01)10841522131479 (Xprezzon), (01)10841522131455 (Qube) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0167-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.