FDA Device recall Z-1793-2026
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
- FDA Device
- Class II
- ongoing
- Published 2026-04-15
On 2026-04-15 the FDA classified a Class II recall of Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor by Spacelabs Healthcare, citing multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measure….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1793-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-02-27 |
| Published | 2026-04-15 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Spacelabs Healthcare |
| Distribution | US |
Product
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Reason
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Identifiers
| code_info | Serial Numbers: UDI-DI: 10841522106415, 10841522106422. 1496-223736, 1496-223737, 1496-223738, 1496-224375, 1496-224376…Serial Numbers: UDI-DI: 10841522106415, 10841522106422. 1496-223736, 1496-223737, 1496-223738, 1496-224375, 1496-224376, 1496-224400, 1496-224401, 1496-224402, 1496-224403, 1496-224404, 1496-224405, 1496-224406, 1496-223850, 1496-223851, 1496-223852, 1496-224448, 1496-224956, 1496-224957, 1496-224958, 1496-224959, 1496-224960, 1496-224961, 1496-224981, 1496-224258, 1496-223595, 1496-223596, 1496-223597, 1496-223598, 1496-223619, 1496-223620, 1496-223621, 1496-223622, 1496-223623, 1496-223624, 1496-224717, 1496-224718, 1496-223854, 1496-224373, 1496-224374, 1496-223894, 1496-223895, 1496-223896, 1496-223897, 1496-223898, 1496-223899, 1496-223900, 1496-223901, 1496-223902, 1496-223903, 1496-223904, 1496-223905, 1496-223906, 1496-223907, 1496-223918, 1496-225043, 1496-225044, 1496-225045, 1496-225046, 1496-225047, 1496-225048, 1496-225049, 1496-225050, 1496-225051, 1496-225052, 1496-225053, 1496-225054, 1496-225055, 1496-225056, 1496-225057, 1496-225058, 1496-225059, 1496-225060, 1496-225 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Other recalls by Spacelabs Healthcare
- Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelab…
FDA Device · Class II · 2026-08-19 · Spacelabs Healthcare - Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A Product…
FDA Device · Class II · 2026-05-13 · Spacelabs Healthcare - Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology In…
FDA Device · Class II · 2025-11-12 · Spacelabs Healthcare - Brand Name: Branding is not applied to this product
FDA Device · Class II · 2025-10-22 · Spacelabs Healthcare - Brand Name: Branding is not applied to this product
FDA Device · Class II · 2025-10-22 · Spacelabs Healthcare
All Spacelabs Healthcare recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1793-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.