RecallRadar

FDA Device recall Z-0964-2026

da Vinci 5, Surgeon Console Column Motor Connector

On 2025-12-31 the FDA classified a Class II recall of da Vinci 5, Surgeon Console Column Motor Connector by Intuitive Surgical, citing surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest,….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0964-2026
ClassificationClass II
Statusongoing
Initiated2024-08-21
Published2025-12-31
Last updated2026-09-13 11:57 UTC
CompanyIntuitive Surgical
DistributionUS

Product

da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

Reason

Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure

Identifiers

code_info
UDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551,…UDI-DI: 00886874119747. Serial Numbers: 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781, 10529917, 10529918, 10529919, 10529920, 10529922, 10534426, 10534428, 10534429, 10534430, 10534752, 10534753, 10534754, 10534755, 10534757, 10543027, 10543028, 10543048, 10543051, 10544670, 10544676, 10544677, 10544678, 10547713, 10547717, 10547724, 10547728, 10551553, 10551554, 10552616, 10552618, 10552620, 10552621, 10556681, 10556682, 10558195, 10558197, 10558198, 10558202, 10560603, 10560604, 10560607, 10562584, 10562585, 10562586, 10562588, 10565609, 10565610, 10565614, 10567141, 10567143, 10567145, 10569365, 10569367, 10569368, 10569369, 10570704, 10570705, 10570706, 10570707, 10573677, 10573678, 10573679, 10573680, 10573681, 10573682, 10573683, 10573684, 10573685, 10573686, 10576019, 10576021,