FDA Device recall Z-2046-2026
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000
- FDA Device
- Class II
- ongoing
- Published 2026-05-13
On 2026-05-13 the FDA classified a Class II recall of Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000 by Intuitive Surgical, citing due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2046-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-04-02 |
| Published | 2026-05-13 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Intuitive Surgical |
| Distribution | US |
Product
Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
Reason
Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
Identifiers
| code_info | Lot Code: UDI: 00886874114216 FDA Medical Device Listing Number: D531324 See attachment Affected Product List UDI:…Lot Code: UDI: 00886874114216 FDA Medical Device Listing Number: D531324 See attachment Affected Product List UDI: 00886874114216 FDA Medical Device Listing Number: D295931 See attachment Affected Product List |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Other recalls by Intuitive Surgical
- da Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-15
FDA Device · Class II · 2026-07-22 · Intuitive Surgical - Da Vinci ASSY, PSS, SP1098, Part Number: 380601
FDA Device · Class II · 2026-07-01 · Intuitive Surgical - Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface:…
FDA Device · Class II · 2026-06-10 · Intuitive Surgical - Intuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-06
FDA Device · Class I · 2026-05-13 · Intuitive Surgical - Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/C…
FDA Device · Class II · 2026-04-08 · Intuitive Surgical
All Intuitive Surgical recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2046-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.