FDA Device recall Z-2521-2026
Da Vinci ASSY, PSS, SP1098, Part Number: 380601
- FDA Device
- Class II
- ongoing
- Published 2026-07-01
On 2026-07-01 the FDA classified a Class II recall of Da Vinci ASSY, PSS, SP1098, Part Number: 380601 by Intuitive Surgical, citing operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engag….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2521-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-04-22 |
| Published | 2026-07-01 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Intuitive Surgical |
| Distribution | US |
Product
Da Vinci ASSY, PSS, SP1098, Part Number: 380601
Reason
Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.
Identifiers
| code_info | UDI: 00886874114605. Serial Numbers: 605408, 606848, 612328, 628399, 609353, 614073, 617794, 616769, 10008062, 635659,…UDI: 00886874114605. Serial Numbers: 605408, 606848, 612328, 628399, 609353, 614073, 617794, 616769, 10008062, 635659, 631694, 638771, 641430, 694998, 645433, 643209, 954097, 656203, 764644, 656319, 946098, 667190, 666480, 655562, 664662, 680530, 679266, 683974, 674171, 676962, 682694, 686777, 701248, 696877, 688725, 688225, 697922, 700316, 698277, 702699, 709536, 701751, 717804, 705366, 704731, 724178, 703152, 722803, 693262, 736468, 736338, 740359, 742366, 732233, 696089, 724835, 716520, 709987, 728391, 751529, 731765, 672743, 766366, 711517, 735003, 735252, 746457, 746623, 718940, 721722, 691190, 744094, 750174, 738718, 10211920, 759253, 761448, 752536, 812651, 754935, 790597, 811147, 793019, 761000, 812650, 919988, 919987, 913011, 913012, 926423, 759791, 926422, 938794, 771653, 938795, 766365, 963841, 797872, 769736, 794437, 970568, 970562, 977045, 10003059, 957781, 973824, 804588, 10009706, 936269, 10003509, 930240, 936268, 800142, 10003510, 982782, 967667, 10009707, 973355, 94458 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2521-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.