FDA Device recall Z-0072-2018
Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia…
- FDA Device
- Class I
- terminated
- Published 2017-11-29
On 2017-11-29 the FDA classified a Class I recall of Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia… by Spacelabs Healthcare, citing arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0072-2018 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2017-10-11 |
| Published | 2017-11-29 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Spacelabs Healthcare |
| Distribution | US |
Product
Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
Reason
Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
Identifiers
| code_info | UDI: 10841522100017 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0072-2018%22