Shimadzu Medical Systems recalls
Shimadzu Medical Systems appears in 26 official U.S. recall notices indexed by RecallRadar, most recently on 2026-06-10.
Company names are normalized from agency data (legal suffixes removed), so related firms may appear under separate pages.
Recall summary
| Recall notices | 26 |
|---|---|
| Published | 2012-11-14 to 2026-06-10 |
| By agency | FDA Device (26) |
| By classification | Class II (26) |
| By status | terminated (15), ongoing (10), completed (1) |
| Distribution named in notices | AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, US, WA, WI, WV |
Recalls
- Digital Angiography System Trinias:UD150G-40 / D150GC-40
FDA Device · Class II · 2026-06-10 · Shimadzu Medical Systems
Reason: Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off. - MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
FDA Device · Class II · 2022-04-13 · Shimadzu Medical Systems
Reason: It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation. - MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
FDA Device · Class II · 2022-04-13 · Shimadzu Medical Systems
Reason: It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation. - MODEL: X-RAY TV SYSTEM SONIALVISION G4
FDA Device · Class II · 2022-04-13 · Shimadzu Medical Systems
Reason: It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation. - Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40
FDA Device · Class II · 2022-03-09 · Shimadzu Medical Systems
Reason: Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode. - Digital Angiography System Bransist safire Generator Model D150GC-40
FDA Device · Class II · 2022-03-09 · Shimadzu Medical Systems
Reason: Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode. - Trinias, Digital Angiography System containing a MH-200, Ceiling Suspended C-arm Support - Product Usage: intended to be used for cardiac a…
FDA Device · Class II · 2020-12-16 · Shimadzu Medical Systems
Reason: The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement le… - Shimadzu Mobile X-Ray System, MobileDaRt Evolution (MX8 Version), with adjustable handle option, Made in Japan - Product Usage: The device…
FDA Device · Class II · 2020-12-16 · Shimadzu Medical Systems
Reason: There is a potential that the adjustable handle option on the X-Ray system may have been installed with improper bolts. This could result in improper driving operation which could lead to harm of individuals around the… - Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designed to provide fluoro…
FDA Device · Class II · 2020-12-16 · Shimadzu Medical Systems
Reason: The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement le… - MH-200S, system with Ceiling Suspended C-arm Support - Product Usage: intended to be used for cardiac angiography, neurovascular angiograph…
FDA Device · Class II · 2020-12-16 · Shimadzu Medical Systems
Reason: The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement le… - Trinias DAP Meter, a component of the Digital Anglo System Trinias contained within the DAR-9500f - Product Usage: It is intended to be use…
FDA Device · Class II · 2020-07-22 · Shimadzu Medical Systems
Reason: The DAP meters were found to be outside of tolerance. - Sonialvision Safire II, Model #: DAR-8000f Product Usage: This angiographic x-ray system device intended to be used for the radiography in…
FDA Device · Class II · 2018-06-20 · Shimadzu Medical Systems
Reason: When selecting serial radiography with a pulse rate of 7.5fps (including selecting a preset or changing the pulse rate using a temporary edit function) it was observed the indicated "irradiation time" on the X-ray Gener… - Shimadzu X-ray TV System, Catalog No. SONIALVISION G4
FDA Device · Class II · 2017-01-04 · Shimadzu Medical Systems
Reason: Shimadzu Medical Systems is recalling Shimadzu X-ray TV System SONIALVISION due to possibility of oil leaking from the x-ray tube housing under certain conditions. - Shimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems with 0.7/1.2JG326D X-Ray Tube Diagnostic x-r…
FDA Device · Class II · 2017-01-04 · Shimadzu Medical Systems
Reason: Shimadzu Medical Systems is recalling the Shimadzu X-ray High Voltage Generator due to the possibility of oil leaking from the x-ray tube housing under certain conditions. - X-Ray Diagnostic Table, Model No. YSF-300 Use in combination with an X-ray high voltage generator, an X-ray tube unit, an Image intensifier…
FDA Device · Class II · 2016-05-04 · Shimadzu Medical Systems
Reason: Report of unintentional movement of table to reverse tilting. - Mobile X-ray system MobileDaRt Evolution/FDR Go Software The device is a mobile X-ray system and allows a technician to take a general digi…
FDA Device · Class II · 2015-12-30 · Shimadzu Medical Systems
Reason: The firm is recalling the Shimadzu Mobile X-ray system because an image may not transfer to image server properly. - Shimadzu C-arm Table, Catalog No. IVS-110, Medical Imaging Table
FDA Device · Class II · 2015-11-18 · Shimadzu Medical Systems
Reason: The C-arm may interfere with the table top under certain circumstances. - Shimadzu Digital Radiography X-Ray System, Catalog No. DAR-8000f
FDA Device · Class II · 2014-05-14 · Shimadzu Medical Systems
Reason: Shimadzu Corporation is recalling the Shimadzu Digital X-Ray System because they have received a report of a lost image under certain circumstances. - Mobile X-ray System, Model: MUX-100, MUX-100H Product Usage: The device is a mobile X-ray system in order to take general radiography of pa…
FDA Device · Class II · 2014-03-12 · Shimadzu Medical Systems
Reason: Shimadzu Corporation is recalling the Shimadzu Mobile X-ray Sytems (MUX-100 and MUX-100H) because the connecting metal plate which connects the main chassis and the column of MUX-100 may crack. - Shimadzu Corporation - Mobile DaRt Evolution This device is a mobile X-ray system in order to take general radiography of patients in the h…
FDA Device · Class II · 2013-06-12 · Shimadzu Medical Systems
Reason: The recall was initiated because Shimadzu Corporation has identified a potential issue in certain manufacturing lots of the Shimadzu MobileDaRt Evolution in combination with CXDI-70C, CXDI-55C, and CXDI-55G. As a result… - 0.7U163CS-36 In combination with below systems MUX-100H, MUX-100D, MobileArt Evolution Brand Name: MUX-100H. Classification Name: Mobile x-…
FDA Device · Class II · 2012-11-14 · Shimadzu Medical Systems
Reason: Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1)… - R-30H , Diagnostic X-Ray Beam-Limiting Device
FDA Device · Class II · 2012-11-14 · Shimadzu Medical Systems
Reason: Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1)… - 0.7/1.2JG326D-265 Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital who c…
FDA Device · Class II · 2012-11-14 · Shimadzu Medical Systems
Reason: Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1)… - R-20J , Diagnostic X-Ray Beam-Limiting Device
FDA Device · Class II · 2012-11-14 · Shimadzu Medical Systems
Reason: Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1)… - 0.4/0.7JG326D-265AX/AT Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital…
FDA Device · Class II · 2012-11-14 · Shimadzu Medical Systems
Reason: Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1)… - 0.7/1.3U163CS-36 In combination with below systems MobileArt Evolution, MobileDaRt Evolution Product Usage: The device is a mobile X-ray sy…
FDA Device · Class II · 2012-11-14 · Shimadzu Medical Systems
Reason: Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1)…