R-30H , Diagnostic X-Ray Beam-Limiting Device

On 2012-11-14 the FDA classified a Class II recall of R-30H , Diagnostic X-Ray Beam-Limiting Device by Shimadzu Medical Systems, citing shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standar….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0260-2013
ClassificationClass II
Statusterminated
Initiated2011-10-28
Published2012-11-14
Last updated2026-09-13 11:53 UTC
CompanyShimadzu Medical Systems
DistributionUS

Product

R-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85, 0.6/1.2P324DK-125, 0.6/1.2P324DK-85, 0.6/1.2P38DE-85 Device Model # R-30H. Subsequent Product Code: IZX Product Usage: The device is the beam-limiting device for X-ray Radiography.

Reason

Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.

Identifiers

code_infoDevice Model #R-30H

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.