FDA Device recall Z-0260-2013
R-30H , Diagnostic X-Ray Beam-Limiting Device
- FDA Device
- Class II
- terminated
- Published 2012-11-14
On 2012-11-14 the FDA classified a Class II recall of R-30H , Diagnostic X-Ray Beam-Limiting Device by Shimadzu Medical Systems, citing shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standar….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0260-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2011-10-28 |
| Published | 2012-11-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Shimadzu Medical Systems |
| Distribution | US |
Product
R-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85, 0.6/1.2P324DK-125, 0.6/1.2P324DK-85, 0.6/1.2P38DE-85 Device Model # R-30H. Subsequent Product Code: IZX Product Usage: The device is the beam-limiting device for X-ray Radiography.
Reason
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
Identifiers
| code_info | Device Model #R-30H |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0260-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.