FDA Device recall Z-0262-2013
0.7U163CS-36 In combination with below systems MUX-100H, MUX-100D, MobileArt Evolution Brand Name: MUX-100H. Classification Name: Mobile x-…
- FDA Device
- Class II
- terminated
- Published 2012-11-14
On 2012-11-14 the FDA classified a Class II recall of 0.7U163CS-36 In combination with below systems MUX-100H, MUX-100D, MobileArt Evolution Brand Name: MUX-100H. Classification Name: Mobile x-… by Shimadzu Medical Systems, citing shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standar….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0262-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2011-10-28 |
| Published | 2012-11-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Shimadzu Medical Systems |
| Distribution | US |
Product
0.7U163CS-36 In combination with below systems MUX-100H, MUX-100D, MobileArt Evolution Brand Name: MUX-100H. Classification Name: Mobile x-ray system. Device Model#MUX-100H. Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.
Reason
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
Identifiers
| code_info | S/N of X-ray tube: 63049 63048 63074 63081 63082 63077 63350 63349 63629 63647 63909 62729 64214 66033 6…S/N of X-ray tube: 63049 63048 63074 63081 63082 63077 63350 63349 63629 63647 63909 62729 64214 66033 66032 70070 70065 70364 66327 66331 70058 66325 70955 71241 70920 71848 72472 72462 73056 73058 73341 73675 74577 74563 74565 74557 74558 74846 74845 75455 75452 75457 75728 75737 75736 76059 76057 76060 76054 76062 76360 76366 76612 76636 76336 76916 76655 76647 76938 76940 2XY080646 2XY0081249 2XY0081245 2XY0081543 2XY0081858 2XY0081849 2XY0081855 2XY0081856 2XY0081865 2XY0081862 2XY0082746 2XY0082764 2XY0083055 2XY0083052 2XY0083064 2XY0083051 2XY0083053 2XY0083038 2XY0083063 2XY0083063 2XY0083065 2XY0083343 2XY0083340 2XY0083349 2XY0083354 2XY0083360 2XY0083355 2XY0083646 2XY0083359 2XY0083652 2XY0083670 2XY0083674 2XY0083655 2XY0083673 2XY0083675 2XY0083665 2XY0083645 2XY0083643 2XY0083629 2XY0074577 2XY0083638 2XY0083943 2XY0083671 2XY0083933 2XY0083930 2XY0083925 2XY008453 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0262-2013%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.