RecallRadar

FDA Device recall Z-3308-2026

Ingenia 1.5T, 1.5T CX, and 1.5T S

On 2026-10-07 the FDA classified a Class II recall of Ingenia 1.5T, 1.5T CX, and 1.5T S by Philips North America, citing when exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3308-2026
ClassificationClass II
Statusongoing
Initiated2026-08-28
Published2026-10-07
Last updated2026-10-13 05:30 UTC
CompanyPhilips North America
DistributionUS

Product

Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and R13;

Reason

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Identifiers

code_info
1. Model number (REF): 1. 781315 (1.5T); UDI-DI: 00884838099715; Serial Numbers: 61005, 61007, 61014, 61033, 61036, 610…1. Model number (REF): 1. 781315 (1.5T); UDI-DI: 00884838099715; Serial Numbers: 61005, 61007, 61014, 61033, 61036, 61047, 61056, 61057, 61076, 61087, 61088, 61105, 61116; 2. Model Number (REF): 781341 (1.5T); UDI-DI: 00884838055322; Serial Numbers: 70042, 70051, 70056, 70062, 70203, 70227, 70229, 70249, 70257, 70262, 70285, 70300, 70311, 70383, 70424, 70432, 70438, 70445, 70453, 70460, 70472, 70588, 70598, 70619, 70638, 70642, 70645, 70647, 70708, 70717, 70724, 70728, 70743, 70750, 70751, 70787, 70801, 70821, 70825, 70836, 70846, 70875, 70876, 70886, 70889, 70896, 70904, 70920, 70957, 70961, 70967, 70970, 70979, 70988, 84049, 84076, 84079, 84080, 84141, 84162, 84170, 84173, 84186, 84205, 84212, 84219, 84233, 84250, 84254, 84266, 84267, 84281, 84289, 84329, 84332, 84335, 84343, 84359, 84360, 84368, 84379, 84388, 84392, 84403, 84415, 84431, 84433, 84436, 84450, 84451, 84467, 84476, 84491, 84500, 84510, 84539, 84549, 84555, 84556, 84568, 84596, 84634, 84635, 84641, 84642, 84652, 84660,

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-13 05:30 UTC. Current status: ongoing.

Other recalls by Philips North America

All Philips North America recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3308-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.