MR 7700

On 2026-10-07 the FDA classified a Class II recall of MR 7700 by Philips North America, citing when exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3313-2026
ClassificationClass II
Statusongoing
Initiated2026-08-28
Published2026-10-07
Last updated2026-10-13 05:30 UTC
CompanyPhilips North America
DistributionUS

Product

MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R13;

Reason

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Identifiers

code_info
1. Model number (REF): 782120; UDI-DI: 00884838104112; Serial Numbers: 65000, 65001, 65002, 65003, 65004, 65005, 65006,…1. Model number (REF): 782120; UDI-DI: 00884838104112; Serial Numbers: 65000, 65001, 65002, 65003, 65004, 65005, 65006, 65008, 65009, 65011, 65012, 65013, 65014, 65015, 65016, 65017, 65018, 65019, 65020, 65021, 65023, 65024, 65025, 65026, 65027, 65028, 65029, 65030, 65031, 65032, 65034, 65035, 65038, 65039, 65040, 65041, 65042, 65043, 65049, 65050, 65051, 65056, 65059, 65060, 65061, 65063, 65064, 65065, 65066, 65067; 2. Model Number (REF): 782153; UDI-DI: 00884838112858; Serial Numbers: 68000, 68001, 68006, 68007, 68008, 68009, 68010, 68013, 68014, 68016, 68017, 68018, 68019, 68020, 68021, 68022, 68023, 68024, 68025, 68026, 68028, 68029, 68030, 68031, 68032, 68033, 68034, 68036, 68037, 68041, 68042, 68043, 68047, 68049, 68050; 3. Model Number (REF): 782155; UDI-DI: 00884838115668; Serial Numbers: 94000, 94002, 94004, 94006, 94007, 94008, 94009, 94011, 94012, 94013, 94014, 94015, 94016, 94021, 94023, 94024; 4. Model Number (REF): 782199; UDI-DI: 00884838128439; Serial Numbers: 19300

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-13 05:30 UTC. Current status: ongoing.

Other recalls by Philips North America

All Philips North America recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3313-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.