RecallRadar

FDA Device recall Z-3307-2026

Ingenia Blue Seal SE and Blue Seal XE

On 2026-10-07 the FDA classified a Class II recall of Ingenia Blue Seal SE and Blue Seal XE by Philips North America, citing when exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3307-2026
ClassificationClass II
Statusongoing
Initiated2026-08-28
Published2026-10-07
Last updated2026-10-13 05:30 UTC
CompanyPhilips North America
DistributionUS

Product

Ingenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13;

Reason

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Identifiers

code_info
1. Model number (REF): 1. 782185 (SE); UDI-DI: 00884838125698; Serial Number: 53002; 2. Model Number (REF): 782192 (X…1. Model number (REF): 1. 782185 (SE); UDI-DI: 00884838125698; Serial Number: 53002; 2. Model Number (REF): 782192 (XE); UDI-DI: 00884838128422; Serial Numbers: 56000, 56001, 56005, 56010, 56012;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-13 05:30 UTC. Current status: ongoing.

Other recalls by Philips North America

All Philips North America recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3307-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.