RecallRadar

FDA Device recall Z-3231-2026

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

On 2026-09-30 the FDA classified a Class II recall of VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811 by Biomerieux, citing the issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3231-2026
ClassificationClass II
Statusongoing
Initiated2026-07-24
Published2026-09-30
Last updated2026-10-06 05:30 UTC
CompanyBiomerieux
Distributionn/a

Product

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Reason

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.

Identifiers

code_infoUDI-DI 03573026637712; Lot Number 1333393504

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-06 05:30 UTC. Current status: ongoing.

Other recalls by Biomerieux

All Biomerieux recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3231-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.