RecallRadar

FDA Device recall Z-1464-2024

VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984

On 2024-05-01 the FDA classified a Class I recall of VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984 by Biomerieux, citing ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is pr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1464-2024
ClassificationClass I
Statusongoing
Initiated2024-03-14
Published2024-05-01
Last updated2026-09-13 11:57 UTC
CompanyBiomerieux
Distributionn/a

Product

VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST-N402, REF 423644; j) AST-N404, REF 423664; k) AST-N802, REF 423706; l) AST-N405, REF 423864; m) AST-N417, REF 423880; n) AST-N408, REF 423924; o) AST-N409, REF 423925; p) AST-N420, REF 424039; q) AST-N423, REF 424042; r) AST-N422, REF 424056; s) AST-N436, REF 424440; t) AST-N809, REF 424703; u) AST-N806, REF 424709; v) AST-N807, REF 424710; w) AST-N808, REF 424711; x) AST-N810, REF 424712; y) AST-N812, REF 424721

Reason

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

Identifiers

code_info
a) AST-N334, REF 418984, UDI/DI 3573026490720, Lot/Serial Numbers: 7742790503, 7742805403, 7742828503, 7742830203; b…a) AST-N334, REF 418984, UDI/DI 3573026490720, Lot/Serial Numbers: 7742790503, 7742805403, 7742828503, 7742830203; b) AST-GN95, REF 421982, UDI/DI 3573026565923, Lot/Serial Numbers: 6852740103, 6852751503, 6852778503, 6852811503; c) AST-GN99, REF 423102, UDI/DI 3573026608132, Lot/Serial Numbers: 6892721103, 6892727403, 6892740503, 6892740403, 6892743503, 6892761503, 6892770403, 6892790103, 6892799403, 6892799103, 6892810103, 6892818103, 6892831503; d) AST-N390, REF 423340, UDI/DI 3573026612177, Lot/Serial Numbers: 1402832104, 1402761504; e) AST-N391, REF 423341, UDI/DI 3573026612269, Lot/Serial Numbers: 1412831104, 1412754504; f) AST-N392, REF 423342, UDI/DI 3573026612283, Lot/Serial Numbers: 1422819204; g) AST-N395, REF 423491, UDI/DI 3573026615802, Lot/Serial Numbers: 1452725204; h) AST-N401, REF 423643, UDI/DI 3573026616816, Lot/Serial Numbers: 1512776404, 1512789204, 1512811104, 1512831104, 1512720404, 1512729404, 1512729104, 1512748504, 1512754104, 151278540