RecallRadar

FDA Device recall Z-3155-2026

Philips Medical Systems Technologies Ltd

On 2026-10-07 the FDA classified a Class II recall of Philips Medical Systems Technologies Ltd by Philips North America, citing the Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3155-2026
ClassificationClass II
Statusongoing
Initiated2026-08-18
Published2026-10-07
Last updated2026-10-13 05:30 UTC
CompanyPhilips North America
Distributionn/a

Product

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Reason

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

Identifiers

code_infoUDI: (01)00884838101111

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-13 05:30 UTC. Current status: ongoing.

Other recalls by Philips North America

All Philips North America recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3155-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.