RecallRadar

FDA Device recall Z-3143-2026

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number…

On 2026-09-23 the FDA classified a Class II recall of Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number… by Ge Medical Systems Information Technologies, citing device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3143-2026
ClassificationClass II
Statusongoing
Initiated2026-07-30
Published2026-09-23
Last updated2026-09-29 05:30 UTC
CompanyGe Medical Systems Information Technologies
DistributionUS

Product

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

Reason

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

Identifiers

code_info
5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial N…5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-29 05:30 UTC. Current status: ongoing.

Other recalls by Ge Medical Systems Information Technologies

All Ge Medical Systems Information Technologies recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3143-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.