FDA Device recall Z-2279-2023
Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA
- FDA Device
- Class II
- ongoing
- Published 2023-08-09
On 2023-08-09 the FDA classified a Class II recall of Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA by Ge Oec Medical Systems, citing image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2279-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-05-15 |
| Published | 2023-08-09 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Ge Oec Medical Systems |
| Distribution | US |
Product
Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems
Reason
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
Identifiers
| code_info | P/N/Sales Order Number: 5451208/80165102, 80171955, 80173514, 80162453, 80183950, 80151733, 80163547, 80171532, 801564…P/N/Sales Order Number: 5451208/80165102, 80171955, 80173514, 80162453, 80183950, 80151733, 80163547, 80171532, 80156410, 80176650, 10530514, 105449333, 80167595; 5485304/10557644, 118687327; 5761613/10283014, 10152211, 105108003, 105114645. OEC 9800, UDI-DI: 00840682114349; OEC 9900 Systems, UDI-DI: 00840682114301 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2279-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.