RecallRadar

FDA Device recall Z-0706-2025

OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile

On 2024-12-25 the FDA classified a Class II recall of OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile by Ge Oec Medical Systems, citing due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0706-2025
ClassificationClass II
Statusongoing
Initiated2024-11-21
Published2024-12-25
Last updated2026-09-13 11:57 UTC
CompanyGe Oec Medical Systems
DistributionAK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Product

OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

Reason

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Identifiers

code_info
Product Name: OEC 9800 GTIN code: 00840682114349 Lot/Serial Numbers / UDI codes: 82XXXX04237 (01)00840682114349(11)1706…Product Name: OEC 9800 GTIN code: 00840682114349 Lot/Serial Numbers / UDI codes: 82XXXX04237 (01)00840682114349(11)170600(21)82XXXX04237 8S-2249 Not Applicable-MFG date prior to UDI compliance requirement 89-1022 Not Applicable-MFG date prior to UDI compliance requirement 82-1772 Not Applicable-MFG date prior to UDI compliance requirement 82-0980 Not Applicable-MFG date prior to UDI compliance requirement 8S-1793 Not Applicable-MFG date prior to UDI compliance requirement 82-2349 Not Applicable-MFG date prior to UDI compliance requirement 82-3361 Not Applicable-MFG date prior to UDI compliance requirement 8S-1398 Not Applicable-MFG date prior to UDI compliance requirement 82-0908 Not Applicable-MFG date prior to UDI compliance requirement 82-0599 Not Applicable-MFG date prior to UDI compliance requirement 8S-0533 Not Applicable-MFG date prior to UDI compliance requirement 82-0189 Not Applicable-MFG date prior to UDI compliance requirement 8S-0186 Not Applicable-MFG date prior to UDI co

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0706-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.