FDA Device recall Z-2160-2021
OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot imag…
- FDA Device
- Class II
- terminated
- Published 2021-08-04
On 2021-08-04 the FDA classified a Class II recall of OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot imag… by Ge Oec Medical Systems, citing there is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2160-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-06-10 |
| Published | 2021-08-04 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Ge Oec Medical Systems |
| Distribution | US |
Product
OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures
Reason
There is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
Identifiers
| code_info | All OEC 3D Systems |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2160-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.