OEC 8800 Flexiview designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional pro…

On 2012-07-25 the FDA classified a Class II recall of OEC 8800 Flexiview designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional pro… by Ge Oec Medical Systems, citing please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imag….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2027-2012
ClassificationClass II
Statusterminated
Initiated2007-11-19
Published2012-07-25
Last updated2026-09-13 11:53 UTC
CompanyGe Oec Medical Systems
Distributionn/a

Product

OEC 8800 Flexiview designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. The OEC 8800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.

Reason

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

Identifiers

code_infoPart numbers: OEC 8800 All serial numbers.