FDA Device recall Z-1963-2026
Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF):…
- FDA Device
- Class II
- ongoing
- Published 2026-05-06
On 2026-05-06 the FDA classified a Class II recall of Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF):… by Philips North America, citing the potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1963-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-04-14 |
| Published | 2026-05-06 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Identifiers
| code_info | 1. Model Number (REF): 781358. UDI: (01)00884838088115(21)45121, (01)00884838088115(21)45148, (01)00884838088115(21)451…1. Model Number (REF): 781358. UDI: (01)00884838088115(21)45121, (01)00884838088115(21)45148, (01)00884838088115(21)45160, (01)00884838088115(21)45438, (01)00884838088115(21)45034, (01)00884838088115(21)45049, (01)00884838088115(21)45066, (01)00884838088115(21)45090, (01)00884838088115(21)45141, (01)00884838088115(21)45149, (01)00884838088115(21)45170, (01)00884838088115(21)45250, (01)00884838088115(21)45290, (01)00884838088115(21)45085, (01)00884838088115(21)45312, (01)00884838088115(21)45376, (01)00884838088115(21)45122, (01)00884838088115(21)45169, (01)00884838088115(21)45191, (01)00884838088115(21)45192, (01)00884838088115(21)45203, (01)00884838088115(21)45204, (01)00884838088115(21)45207, (01)00884838088115(21)45220, (01)00884838088115(21)45228, (01)00884838088115(21)45229, (01)00884838088115(21)45249, (01)00884838088115(21)45278, (01)00884838088115(21)45279, (01)00884838088115(21)45320, (01)00884838088115(21)45028, (01)00884838088115(21)45195, (01)00884838088115(21)45174, (01)008 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1963-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.