RecallRadar

FDA Device recall Z-1901-2018

Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz

On 2018-05-30 the FDA classified a Class II recall of Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz by Livanova Usa, citing firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1901-2018
ClassificationClass II
Statusterminated
Initiated2018-04-20
Published2018-05-30
Last updated2026-09-13 11:54 UTC
CompanyLivanova Usa
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WI, WY

Product

Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz

Reason

Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.

Identifiers

code_infoSerial #: a) 16S10743 - 6S11708 b) 16S10772 - 16S16840 C) 16S10958 - 16S16847

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1901-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.