FDA Device recall Z-1073-2025
SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation
- FDA Device
- Class II
- ongoing
- Published 2025-02-12
On 2025-02-12 the FDA classified a Class II recall of SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation by Livanova Usa, citing vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1073-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-12-12 |
| Published | 2025-02-12 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Livanova Usa |
| Distribution | AK, AR, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV |
Product
SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).
Reason
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Identifiers
| code_info | UDI-DI: 05425025750405. All Serial Numbers less than 500,000. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Livanova Usa
- SenTiva, REF: 1000
FDA Device · Class II · 2025-02-12 · Livanova Usa - Brand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: T…
FDA Device · Class II · 2025-02-05 · Livanova Usa - Brand Name: SenTiva DUO" Product Name: VNS Therapy¿ SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software Version: NA Product De…
FDA Device · Class II · 2025-02-05 · Livanova Usa - Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303
FDA Device · Class II · 2022-02-09 · Livanova Usa - VNS Therapy SENTIVA DUO # 1000-D - Product Usage: indicated for the treatment of chronic or recurrent depression in patients who are experi…
FDA Device · Class II · 2020-12-16 · Livanova Usa
Source