FDA Device recall Z-1725-2014
Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes
- FDA Device
- Class II
- terminated
- Published 2014-06-18
On 2014-06-18 the FDA classified a Class II recall of Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes by Leica Microsystems, citing the ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down a….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1725-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-09-11 |
| Published | 2014-06-18 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Leica Microsystems |
| Distribution | AZ, IL, MA, MO, NC, OR, TX, VA, WI |
Product
Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.
Reason
The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).
Identifiers
| code_info | Leica Ceiling Mount MSF-1; Article Numbers: 10446781, 10446782 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1725-2014%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.