RecallRadar

FDA Device recall Z-1725-2014

Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes

On 2014-06-18 the FDA classified a Class II recall of Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes by Leica Microsystems, citing the ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1725-2014
ClassificationClass II
Statusterminated
Initiated2013-09-11
Published2014-06-18
Last updated2026-09-13 11:53 UTC
CompanyLeica Microsystems
DistributionAZ, IL, MA, MO, NC, OR, TX, VA, WI

Product

Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Reason

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

Identifiers

code_infoLeica Ceiling Mount MSF-1; Article Numbers: 10446781, 10446782