RecallRadar

FDA Device recall Z-0477-2023

Leica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope system

On 2022-12-21 the FDA classified a Class II recall of Leica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope system by Leica Microsystems, citing there is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0477-2023
ClassificationClass II
Statusongoing
Initiated2022-11-21
Published2022-12-21
Last updated2026-09-13 11:56 UTC
CompanyLeica Microsystems
DistributionUS

Product

Leica MICROSYSTEMS M530 OHX, Part Number 10448737, surgical microscope system

Reason

There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.

Identifiers

code_info
Serial Numbers: 10622001, 10622002, 10822001, 20322002, 20522001, 20622001, 30122001, 30322001, 30522001, 30622002, 307…Serial Numbers: 10622001, 10622002, 10822001, 20322002, 20522001, 20622001, 30122001, 30322001, 30522001, 30622002, 30722001, 31221001, 40422001, 41221001, 50122001, 60122001, 60522001, 60622001, 70122001, 70322001, 70422001, 70622001, 80422001, 80422002, 80722001, 80822001, 90122001, 90622001, 91221001, 100122001, 100122002, 100322002, 100622001, 100622002, 110222001, 110522001, 110822001, 111221001, 120222001, 120522001, 130122001, 130522001, 130622001, 140122001, 140222001, 140422001, 140522001, 141221001, 151221001, 160122001, 160322001, 160622001, 160622002, 170222001, 170522001, 190322001, 190622001, 200122001, 200522001, 200722001, 210422001, 210622001, 230222001, 230622001, 240122001, 240522001, 260522001, 260722001, 260722002, 261121001, 261221001, 270122001, 270422001, 270422002, 270522001, 270622001, 271121001, 280222002, 280622001, 290522001, 290622001, 291221001, 300322001, 300622002, 301121001, 301221001, 310322001, 310522001. ***EXPANSION 09/08/2023, Serial Numbers: 2

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0477-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.