RecallRadar

FDA Device recall Z-1721-2025

Washer Disinfector Aquadis 56

On 2025-05-14 the FDA classified a Class II recall of Washer Disinfector Aquadis 56 by Getinge Disinfection Ab, citing the frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1721-2025
ClassificationClass II
Statusongoing
Initiated2025-04-02
Published2025-05-14
Last updated2026-09-13 11:57 UTC
CompanyGetinge Disinfection Ab
DistributionUS

Product

Washer Disinfector Aquadis 56; Model Number: 56M

Reason

The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.

Identifiers

code_info
Model: 56M; UDI-DI: 07340153710351; Serial Numbers: WAA105199, WAA105251, WAA105268, WAA105357, WAA105535, WAA105700, W…Model: 56M; UDI-DI: 07340153710351; Serial Numbers: WAA105199, WAA105251, WAA105268, WAA105357, WAA105535, WAA105700, WAA105915, WAA105972, WAA105973, WAA106079, WAA106114, WAA106135, WAA106451, WAA106471, WAA106487, WAA106505, WAA107436, WAA107492, WAA107565, WAA107581, WAA107588, WAA107920, WAA108225, WAA108807, WAA108920, WAA108921, WAA108922, WAA109036, WAA109047, WAA109231, WAA109377, WAA109392, WAA109635, WAA109687, WAA109731, WAA109790, WAA109835, WAA109840, WAA109853, WAA109907, WAA109972, WAA110109, WAA110366, WAA110487, WAA110546, WAA110557, WAA110711, WAA110726, WAA110802, WAA110847, WAA110862, WAA110984, WAA110999, WAA111007, WAA111011, WAA111068, WAA111086, WAA111087, WAA111200, WAA600033, WAA600054, WAA600100, WAA600102, WAA600103, WAA600145, WAA600190, WAA600282, WAA600369, WAA600431, WAA600434, WAA600444, WAA600445, WAA600553, WAA600554, WAA600587, WAA600588, WAA600609, WAA600610, WAA600704, WAA106987, WAA110250, WAA110257, WAA108634, WAA110823, WAA108206, WAA109799, WA

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1721-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.