FDA Device recall Z-0016-2026

Getinge 88-Series Washer- Disinfector

On 2025-10-08 the FDA classified a Class II recall of Getinge 88-Series Washer- Disinfector by Getinge Disinfection Ab, citing potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not tr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0016-2026
ClassificationClass II
Statusongoing
Initiated2025-08-29
Published2025-10-08
Last updated2026-09-13 11:57 UTC
CompanyGetinge Disinfection Ab
DistributionUS

Product

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Reason

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Identifiers

code_infoModel Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.