RecallRadar

FDA Device recall Z-1696-2024

OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use

On 2024-05-08 the FDA classified a Class II recall of OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use by Advanced Instruments, citing system error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1696-2024
ClassificationClass II
Statusongoing
Initiated2024-03-11
Published2024-05-08
Last updated2026-09-13 11:57 UTC
CompanyAdvanced Instruments
DistributionUS

Product

OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0

Reason

System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.

Identifiers

code_info
UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 2…UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 23040284A, 23020159A, 23050410A, 23030236A, 23020165A, 23020156A, 23020161A, 23030239A, 23030210A, 23030233A, 23030232A, 23020158A, 23030206A, 22111360A, 23020155A, 23030209A, 23040287A, 22111362A, 23020160A, 23040288A, 23010052A, 22111354A, 23010055A, 22111350A, 22111365A, 23010053A, 22111363A, 23030237A, 23010062A, 23010015A, 23030202A, 23030207A, 23040290A, 23030204A, 23020157A, 22111367A, 23020158A, 23030203A, 23030240A, 22111358A, 22111359A, 23040286A, 22111355A, 23030205A, 23020164A, 22111368A, 23030238A, 23010060A, 23010061A, 23020163A, 23010011A22111364A. All serial numbers that end in A

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1696-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.