FDA Device recall Z-1696-2024
OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use
- FDA Device
- Class II
- ongoing
- Published 2024-05-08
On 2024-05-08 the FDA classified a Class II recall of OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use by Advanced Instruments, citing system error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1696-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-03-11 |
| Published | 2024-05-08 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Advanced Instruments |
| Distribution | US |
Product
OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0
Reason
System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.
Identifiers
| code_info | UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 2…UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 23040284A, 23020159A, 23050410A, 23030236A, 23020165A, 23020156A, 23020161A, 23030239A, 23030210A, 23030233A, 23030232A, 23020158A, 23030206A, 22111360A, 23020155A, 23030209A, 23040287A, 22111362A, 23020160A, 23040288A, 23010052A, 22111354A, 23010055A, 22111350A, 22111365A, 23010053A, 22111363A, 23030237A, 23010062A, 23010015A, 23030202A, 23030207A, 23040290A, 23030204A, 23020157A, 22111367A, 23020158A, 23030203A, 23030240A, 22111358A, 22111359A, 23040286A, 22111355A, 23030205A, 23020164A, 22111368A, 23030238A, 23010060A, 23010061A, 23020163A, 23010011A22111364A. All serial numbers that end in A |
|---|