Advanced Instruments recalls
Advanced Instruments appears in 2 official U.S. recall notices indexed by RecallRadar, most recently on 2024-05-08.
Company names are normalized from agency data (legal suffixes removed), so related firms may appear under separate pages.
Recall summary
| Recall notices | 2 |
|---|---|
| Published | 2018-12-26 to 2024-05-08 |
| By agency | FDA Device (2) |
| By classification | Class II (2) |
| By status | ongoing (1), terminated (1) |
| Distribution named in notices | US |
Recalls
- OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use
FDA Device · Class II · 2024-05-08 · Advanced Instruments
Reason: System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls. - GloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter System is intended for use by trained healthcare professionals in c…
FDA Device · Class II · 2018-12-26 · Advanced Instruments
Reason: RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will result in failed QC check, preventing user from testing patient samples for RBC.