FDA Device recall Z-1685-2025
epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System
- FDA Device
- Class II
- ongoing
- Published 2025-05-07
On 2025-05-07 the FDA classified a Class II recall of epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System by Siemens Healthcare Diagnostics, citing the occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1685-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-03-19 |
| Published | 2025-05-07 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
Reason
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Identifiers
| code_info | Siemens Material Number (SMN): 10736382; Udi-DI: 00809708072254; Lot Numbers: All lot numbers with prefix (xx) 00 or 06…Siemens Material Number (SMN): 10736382; Udi-DI: 00809708072254; Lot Numbers: All lot numbers with prefix (xx) 00 or 06; epoc Software Version and Senor Configuration: epoc Host SW V3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Siemens Healthcare Diagnostics
- Atellica CH Revised C-Reactive Protein (RCRP)
FDA Device · Class II · 2026-10-07 · Siemens Healthcare Diagnostics - Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
FDA Device · Class II · 2026-08-26 · Siemens Healthcare Diagnostics - Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL
FDA Device · Class II · 2026-08-26 · Siemens Healthcare Diagnostics - Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL
FDA Device · Class III · 2026-08-12 · Siemens Healthcare Diagnostics - Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic
FDA Device · Class II · 2026-08-12 · Siemens Healthcare Diagnostics
All Siemens Healthcare Diagnostics recalls