FDA Device recall Z-2976-2026
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
- FDA Device
- Class II
- ongoing
- Published 2026-08-26
On 2026-08-26 the FDA classified a Class II recall of Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL by Siemens Healthcare Diagnostics, citing due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2976-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-07-13 |
| Published | 2026-08-26 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Reason
Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
Identifiers
| code_info | Lots: 2597, 2627, 2709/UDI: 00842768023293 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.