Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-13 Software Version: All software versions Product Descri…

On 2025-04-30 the FDA classified a Class II recall of Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-13 Software Version: All software versions Product Descri… by Abbott Laboratories, citing when expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1655-2025
ClassificationClass II
Statusongoing
Initiated2025-03-18
Published2025-04-30
Last updated2026-09-13 11:57 UTC
CompanyAbbott Laboratories
DistributionUS

Product

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-13 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No

Reason

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Identifiers

code_infoUDI-DI 00380740017170 SN: 34036BG96