RecallRadar

FDA Device recall Z-0961-2026

Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20

On 2025-12-24 the FDA classified a Class II recall of Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20 by Abbott Laboratories, citing abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0961-2026
ClassificationClass II
Statusongoing
Initiated2025-11-13
Published2025-12-24
Last updated2026-09-13 11:57 UTC
CompanyAbbott Laboratories
DistributionUS

Product

Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20

Reason

Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance issues found in the Alinity ci-series System software versions 3.6.1. and lower that could lead to erroneous results for multiple analytes. The issue was identified by Abbott during the internal testing of complaint investigations.

Identifiers

code_info
UDI-DI: 00380740137410. GTIN: 06952999490048. Software versions 3.6.1 and lower. Serial Numbers: SCM01010 SCM01012 SCM…UDI-DI: 00380740137410. GTIN: 06952999490048. Software versions 3.6.1 and lower. Serial Numbers: SCM01010 SCM01012 SCM01016 SCM01035 SCM01043 SCM01051 SCM01063 SCM01069 SCM01075 SCM01076 SCM01078 SCM01082 SCM01085 SCM01086 SCM01087 SCM01088 SCM01092 SCM01093 SCM01094 SCM01095 SCM01098 SCM01099 SCM01101 SCM01102 SCM01106 SCM01107 SCM01110 SCM01115 SCM01116 SCM01120 SCM01121 SCM01122 SCM01123 SCM01124 SCM01126 SCM01127 SCM01128 SCM01129 SCM01132 SCM01134 SCM01135 SCM01136 SCM01137 SCM01138 SCM01142 SCM01143 SCM01149 SCM01150 SCM01151 SCM01154 SCM01163 SCM01166 SCM01170 SCM01171 SCM01172 SCM01174 SCM01175 SCM01176 SCM01179 SCM01180 SCM01181 SCM01184 SCM01192 SCM01193 SCM01194 SCM01196 SCM01197 SCM01198 SCM01199 SCM01200 SCM01201 SCM01203 SCM01204 SCM01205 SCM01207 SCM01211 SCM01213 SCM01216 SCM01217 SCM01219 SCM01220 SCM01221 SCM01222 SCM01223 SCM01224 SCM01225 SCM01226 SCM01229 SCM01230 SCM01232 SCM01236 SCM01237 SCM01238 SCM01239 SCM01240 SCM01244 SCM01247 SCM01249 SCM01251 SCM01252 SC

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0961-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.