RecallRadar

FDA Device recall Z-1635-2015

Stryker Orthopaedics Triathlon femoral component

On 2015-05-27 the FDA classified a Class II recall of Stryker Orthopaedics Triathlon femoral component by Stryker Howmedica Osteonics, citing the peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile bar….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1635-2015
ClassificationClass II
Statusterminated
Initiated2014-08-15
Published2015-05-27
Last updated2026-09-13 11:53 UTC
CompanyStryker Howmedica Osteonics
DistributionUS

Product

Stryker Orthopaedics Triathlon femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total reconstructive replacement of the knee joint

Reason

The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

Identifiers

code_info
Item number Lot number 5512-F-201 Lots JYET, KFVZ, KGLX ; 5512-F-202 Lots IPPOP, JYKX, KHZL, KJDN, KJHI, KJKY; 5512-F…Item number Lot number 5512-F-201 Lots JYET, KFVZ, KGLX ; 5512-F-202 Lots IPPOP, JYKX, KHZL, KJDN, KJHI, KJKY; 5512-F-301 Lots JTJA, KKAE, KKNB; 5512-F-302 Lots JXXF, KEYD, KFBY, KFTZ, KFXT, KHAB; 5512-F-401 Lots JWGY, KIED; 5512-F-402 Lots IUAJ, JJHT, JOMB, JWPF, JXXM, KGPE, KJMO, KLUP; 5512-F-501 Lots JXMV, KGKP; 5512-F-502 Lots KHWY, KJBK; 5512-F-601 Lots JIML, JIRX, KFAS, KIIY; 5512-F-602 Lots KEXE, KFOS, KFPI, KJAG, KKJW; 5512-F-701 Lot JKTF; 5512-F-801 Lots JWNL, JWPY; 5515-F-301 Lots IVXZD, JEPOD, JHZRA, JITNA, JITND,JJAFD, JJNYA, JJNYD, JKEND, JSNRD, JXBXA, JYFTD, JYGID, JYMTD,JYPJD, KFWHA, KFXID, KGAXD, KGJTD, KGXNA, KIBTA, KIEBA, KIGRD, KIYUD, KJABD, KJISD, KJRID; 5515-F-302 Lots IOLSD, IOLTD, IPRVD, IPTFD, IVGVA, IVLVD, JGIXD, JIGSD, JIYVD, JJSYA, JJWNA, JNVTA, JOWJD, JPFDA, JPHRA, JPIWA, JSZGD, JVYDD, JWPRA, JXBKD, JXOBA, JXXZD, JYBRA, JYBRD, JYKFD, JYKTA, JYMXA, JYNIA, JYNID, JYUED, JZDWD, JZEHD, JZFDD, KFNEA, KFOBA, KFOBD, KFOHD, KFVND, KGMFA, KGMFD, KGWGA, KGZFA, KH

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1635-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.