RecallRadar

FDA Device recall Z-2160-2017

Hoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC (01)07613327131475(10)V07241 Product Usage: Intended f…

On 2017-05-31 the FDA classified a Class II recall of Hoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC (01)07613327131475(10)V07241 Product Usage: Intended f… by Stryker Howmedica Osteonics, citing stryker is recalling Hoffmann LRF (Limb Reconstruction Frame) System Safety Clip - Bone Transport Strut because there is a potential for loosening of the safety clips from the int….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2160-2017
ClassificationClass II
Statusterminated
Initiated2017-04-26
Published2017-05-31
Last updated2026-09-13 11:54 UTC
CompanyStryker Howmedica Osteonics
DistributionUS

Product

Hoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC (01)07613327131475(10)V07241 Product Usage: Intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, and bone and soft tissue reconstruction in pediatric patients and adults.

Reason

Stryker is recalling Hoffmann LRF (Limb Reconstruction Frame) System Safety Clip - Bone Transport Strut because there is a potential for loosening of the safety clips from the intended position on the Quick Release mechanism of the Transport Struts.

Identifiers

code_infoLot Numbers: J49236, V07241, V30655, V47575 and V49404

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2160-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.