FDA Device recall Z-1618-2024
20" Lead Wires, Replacement Part Number: 1067724-2
- FDA Device
- Class II
- ongoing
- Published 2024-05-01
On 2024-05-01 the FDA classified a Class II recall of 20" Lead Wires, Replacement Part Number: 1067724-2 by Ebi, citing damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, an….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1618-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-03-22 |
| Published | 2024-05-01 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Ebi |
| Distribution | US |
Product
20" Lead Wires, Replacement Part Number: 1067724-2
Reason
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
Identifiers
| code_info | UDI-DI: 00812301020300 All devices distributed since May 1, 2023 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Ebi
- Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to…
FDA Device · Class II · 2024-05-01 · Ebi - Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acquired secondary to tra…
FDA Device · Class II · 2024-05-01 · Ebi - 48" Lead Wires, Replacement Part Number: 1067724-4
FDA Device · Class II · 2024-05-01 · Ebi - Cypher MIS Screw System-The system includes screws, various types and sizes of rods, and set screws
FDA Device · Class II · 2015-01-14 · Ebi - Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical…
FDA Device · Class II · 2014-12-31 · Ebi
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1618-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.