FDA Device recall Z-1615-2021
Medtronic REF: MMT-7020D4 Guardian Sensor (3) for the following packaging: GS3 1PK 2L CANADA/ GTIN: 00763000179595¿ Product Usage: intended…
- FDA Device
- Class II
- ongoing
- Published 2021-05-26
On 2021-05-26 the FDA classified a Class II recall of Medtronic REF: MMT-7020D4 Guardian Sensor (3) for the following packaging: GS3 1PK 2L CANADA/ GTIN: 00763000179595¿ Product Usage: intended… by Medtronic, citing use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1615-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-02-26 |
| Published | 2021-05-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Medtronic |
| Distribution | n/a |
Product
Medtronic REF: MMT-7020D4 Guardian Sensor (3) for the following packaging: GS3 1PK 2L CANADA/ GTIN: 00763000179595¿ Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.
Reason
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Identifiers
| code_info | All Lot numbers within active shelf life |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Other recalls by Medtronic
- Medtronic Sphere-9 Catheter, Model Number AFR-00001
FDA Device · Class II · 2026-05-06 · Medtronic - Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14…
FDA Device · Class II · 2026-01-21 · Medtronic - Aurora EV-ICD and Clinical EV-ICD, single chamber, extravascular implantable cardioverter defibrillators (ICD), Product Number DVEA3E4
FDA Device · Class II · 2025-12-10 · Medtronic - Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Softw…
FDA Device · Class II · 2025-07-30 · Medtronic - Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Sof…
FDA Device · Class II · 2025-07-30 · Medtronic
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1615-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.