FDA Device recall Z-1948-2026
Medtronic Sphere-9 Catheter, Model Number AFR-00001
- FDA Device
- Class II
- ongoing
- Published 2026-05-06
On 2026-05-06 the FDA classified a Class II recall of Medtronic Sphere-9 Catheter, Model Number AFR-00001 by Medtronic, citing ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrilla….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1948-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-03-11 |
| Published | 2026-05-06 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medtronic |
| Distribution | n/a |
Product
Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.
Reason
Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
Identifiers
| code_info | US: Basic UDI Number 0763000B000262284, GTIN: 00763000871741; 00763000973391; 00763000973506; 00763000993269; 0019915…US: Basic UDI Number 0763000B000262284, GTIN: 00763000871741; 00763000973391; 00763000973506; 00763000993269; 00199150003709. OUS: 00763000871734; 00763000973384; 00763000973513; 00763000811341; 00763000911515; 00763000963675; 00199150003716. Serial Numbers: 227076469, 2301234000, 2301321000, 2301373000, 2301543000, 2301547000, 2301551000, 2301552000, 2301556000, 2301558000, 2301559000, 2301560000, 2301709000, 0012573781, 0012576276, 0012578858, 0012583026, 0012583027, 0012590913, 0012590914, 0012594540, 0012594541, 0012609253, 0012609255, 0012609258, 0012614806, 0012614807, 0012614808, 0012619893, 0012619896, 0012619898, 0012619899, 0012623987, 0012623988, 0012623991, 0012631219, 0012631220, 0012631221, 0012631222, 0012635279, 0012635281, 0012635282, 0012635283, 0012635284, 0012641293, 0012641294, 0012641295, 0012641297, 0012643392, 0012643393, 0012643396, 0012643399, 0012645205, 0012645206, 0012645208, 0012655005, 0012655006, 0012655008, 0012655009, 0012655010, 0012655011, |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1948-2026%22