FDA Device recall Z-1602-2021
Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000…
- FDA Device
- Class II
- ongoing
- Published 2021-05-26
On 2021-05-26 the FDA classified a Class II recall of Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000… by Medtronic, citing use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1602-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-02-26 |
| Published | 2021-05-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Medtronic |
| Distribution | n/a |
Product
Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Reason
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Identifiers
| code_info | All Lot numbers within active shelf life |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1602-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.