HeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System

On 2022-06-29 the FDA classified a Class I recall of HeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System by Medtronic, citing batteries for the HVAD system have a weld defect affecting internal components within the HVAD battery, resulting in the battery to malfunction and no longer provide power or prev….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1266-2022
ClassificationClass I
Statusongoing
Initiated2022-05-05
Published2022-06-29
Last updated2026-09-13 11:56 UTC
CompanyMedtronic
DistributionUS

Product

HeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System.

Reason

Batteries for the HVAD system have a weld defect affecting internal components within the HVAD battery, resulting in the battery to malfunction and no longer provide power or prevent the battery from holding a complete charge or properly recharging.

Identifiers

code_info
GTIN 00888707008778. Serial numbers BAT935012, BAT935059, BAT935008, BAT935009, BAT935018, BAT935020, BAT935022, BAT93…GTIN 00888707008778. Serial numbers BAT935012, BAT935059, BAT935008, BAT935009, BAT935018, BAT935020, BAT935022, BAT935023, BAT935047, BAT934999, BAT935003, BAT935010, BAT935329, BAT934928, BAT934934, BAT935160, BAT935078, BAT934997, BAT935027, BAT935048, BAT934953, BAT934933, BAT934948, BAT934853, BAT934854, BAT934858, BAT934895, BAT934899, BAT934901, BAT934903, BAT934904, BAT934905, BAT934927, BAT934932, BAT934935, BAT934940, BAT934979, BAT934981, BAT935109, BAT935112, BAT935116, BAT935117, BAT935120, BAT935122, BAT935124, BAT935127, BAT935132, BAT935133, BAT935134, BAT935136, BAT934846, BAT934848, BAT934849, BAT934850, BAT935137, BAT935142, BAT935143, BAT935145, BAT935146, BAT935148, BAT935150, BAT935151, BAT935152, BAT935154, BAT935155, BAT935156, BAT935157, BAT935158, BAT935161, BAT935162, BAT935163, BAT935164, BAT935260, BAT935301, BAT935302, BAT935303, BAT935201, BAT935202, BAT935203, BAT935172, BAT935173, BAT935174, BAT935175, BAT935176, BAT935177, BAT935178, BAT935179, BAT935