FDA Device recall Z-1235-2026

CellMek SPS Sample Preparation System, REF: C44603, with software

On 2026-02-11 the FDA classified a Class II recall of CellMek SPS Sample Preparation System, REF: C44603, with software by Beckman Coulter, citing sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which cou….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1235-2026
ClassificationClass II
Statusongoing
Initiated2025-12-23
Published2026-02-11
Last updated2026-09-13 11:58 UTC
CompanyBeckman Coulter
DistributionUS

Product

CellMek SPS Sample Preparation System, REF: C44603, with software

Reason

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Identifiers

code_infoUDI-DI: 15099590750312. Software Version 2.3.90.0