RecallRadar

FDA Device recall Z-1099-2019

SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung…

On 2019-04-17 the FDA classified a Class II recall of SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung… by Livanova Usa, citing for six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-1099-2019
ClassificationClass II
Statusterminated
Initiated2018-09-14
Published2019-04-17
Last updated2026-09-13 11:55 UTC
CompanyLivanova Usa
DistributionIN, MO, OH

Product

SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.

Reason

For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).

Identifiers

code_infoSerial numbers 16S13187, 16S14887, 16S15335, 16S15336, 16S15337, and 16S11877.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1099-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.