FDA Device recall Z-0887-2026
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T
- FDA Device
- Class II
- ongoing
- Published 2025-12-17
On 2025-12-17 the FDA classified a Class II recall of Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T by Bard Peripheral Vascular, citing acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sha….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0887-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-11-06 |
| Published | 2025-12-17 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Bard Peripheral Vascular |
| Distribution | US |
Product
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Reason
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
Identifiers
| code_info | UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130 |
|---|