FDA Device recall Z-1244-2025
Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236
- FDA Device
- Class I
- ongoing
- Published 2025-03-12
On 2025-03-12 the FDA classified a Class I recall of Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236 by Bard Peripheral Vascular, citing atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1244-2025 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2025-02-05 |
| Published | 2025-03-12 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Bard Peripheral Vascular |
| Distribution | US |
Product
Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239
Reason
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.
Identifiers
| code_info | Instructions for Use versions prior to ZE10895 revision C1 09/24. REF/UDI-DI/Lot(Expiration): 80236/07640142811855/23…Instructions for Use versions prior to ZE10895 revision C1 09/24. REF/UDI-DI/Lot(Expiration): 80236/07640142811855/230893(2025-12-15), 231100(2026-06-09), 231172(2026-06-09), 231171(2026-06-09), 231526(2026-08-21), 231875(2026-08-21), 240018(2026-11-22), 240017(2026-11-22), 240375(2027-01-18), 240567(2027-01-18), 241075(2024-05-22), 241335(2024-06-19), 241550(2024-07-19), 241777(2024-08-19), 242691(2024-12-13); 80237/07640142811862/230897(2026-04-11), 230898(2026-04-11), 231033(2026-05-31), 231101(2026-06-13), 231170(2026-06-13), 231229(2026-07-07), 231400(2026-07-14), 231414(2026-07-31), 231412(2026-06-30), 231413(2026-07-31), 231411(2026-08-03), 231521(2026-09-01), 231520(2026-07-31), 231522(2026-08-31), 231523(2026-09-01), 231524(2026-09-01), 231525(2026-09-01), 231715(2026-09-22), 231885(2026-09-22), 231884(2026-10-27), 240020(2026-11-22), 240396(2026-12-08), 240530(2027-01-10), 240531(2027-01-10), 240566(2027-01-10), 240568(2027-01-25), 240754(2027-02-06), 240769(2027-02-09), 2 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Bard Peripheral Vascular
- EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with soft…
FDA Device · Class II · 2026-10-07 · Bard Peripheral Vascular - Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 60060
FDA Device · Class II · 2026-02-25 · Bard Peripheral Vascular - Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T
FDA Device · Class II · 2025-12-17 · Bard Peripheral Vascular - Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation…
FDA Device · Class II · 2025-10-08 · Bard Peripheral Vascular - Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210
FDA Device · Class II · 2024-11-27 · Bard Peripheral Vascular
All Bard Peripheral Vascular recalls