HeartWare HVAD Pump Kit, REF 1205

On 2022-12-28 the FDA classified a Class II recall of HeartWare HVAD Pump Kit, REF 1205 by Heartware, citing over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driv….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0749-2023
ClassificationClass II
Statusongoing
Initiated2022-11-29
Published2022-12-28
Last updated2026-09-13 11:56 UTC
CompanyHeartware
Distributionn/a

Product

HeartWare HVAD Pump Kit, REF 1205

Reason

Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.

Identifiers

code_info
Serial Numbers: HW1412, HW1413, HW1416, HW1433, HW1435, HW1490, HW1493, HW1495, HW1496, HW1593, HW1599, HW1641, HW1661…Serial Numbers: HW1412, HW1413, HW1416, HW1433, HW1435, HW1490, HW1493, HW1495, HW1496, HW1593, HW1599, HW1641, HW1661, HW1662, HW1663, HW1664, HW1695, HW1696, HW1698, HW1699, HW1706, HW1708, HW1709, HW1718, HW1727, HW1730, HW1740, HW1741, HW1747, HW1754, HW1760, HW1764, HW1765, HW1768, HW1773, HW1774, HW1777, HW1778, HW1783, HW1785, HW1796, HW1804, HW1808, HW1810, HW1811, HW1815, HW1819, HW1821, HW1822, HW1823, HW1824, HW1825, HW1826, HW1828, HW1830, HW1831, HW1832, HW1839, HW1841, HW1856, HW1857, HW1876, HW1879, HW1882, HW1884, hw1895, HW1896, HW1897, HW1899, HW1900, HW1901, HW1905, HW1906, HW1907, HW1908, HW1913, hw1914, HW1917, HW1920, HW1921, HW1922, HW1923, HW1924, HW1925, HW1926, HW1927, HW1928, HW1929, HW1930, HW1931, HW1933, HW1934, HW1935, HW1936, HW1937, HW1938, HW1939, HW1940, HW1941, HW1942, HW1943, hw1944, HW1947, HW1950, HW1952, HW1960, HW1961, HW1962, HW1964, HW1965, HW1966, HW1967, HW1968, HW1969, HW1970, HW1971, HW1972, HW1973, HW1974, HW1976, HW1977, HW1978, HW1979,