FDA Device recall Z-0481-2024
PUMP 1104 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT EN JA, Model Number 1104JP
- FDA Device
- Class II
- ongoing
- Published 2023-12-13
On 2023-12-13 the FDA classified a Class II recall of PUMP 1104 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT EN JA, Model Number 1104JP by Heartware, citing updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0481-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-10-31 |
| Published | 2023-12-13 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Heartware |
| Distribution | n/a |
Product
PUMP 1104 HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT EN JA, Model Number 1104JP
Reason
Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components
Identifiers
| code_info | GTIN none, Serial Numbers: HW25172, HW25169FSCA, HW25170FSCA, HW25174FSCA; GTIN 00763000170967, Serial numbers: HW4…GTIN none, Serial Numbers: HW25172, HW25169FSCA, HW25170FSCA, HW25174FSCA; GTIN 00763000170967, Serial numbers: HW42722, HW42771, HW42285, HW41696, HW41675, HW40953, HW41848, HW41847, HW41316, HW41828, HW41824, HW41803, HW41670, HW41271, HW41701, HW41674, HW41791, HW41792, HW41804, HW41252, HW41256, HW41700, HW41699, HW41783, HW41317, HW41309, HW41673, HW41254, HW40651, HW40665, HW40663, HW41672, HW41308, HW41269, HW41544, HW41270, HW41214, HW41314, HW40417, HW41263, HW41264, HW41265, HW41047, HW41260, HW41266, HW41251, HW41259, HW41258, HW40579, HW41048, HW41262, HW41045, HW40961, HW40958, HW40956, HW41050, HW41255, HW40881, HW40789, HW40957, HW40960, HW40951, HW40659, HW40602, HW40816, HW40954, HW40950, HW40662, HW40664, HW40580, HW40668, HW40613, HW40652, HW40666, HW40667, HW40660, HW40658, HW40612, HW40420, HW40657, HW40607, HW40552, HW40609, HW40599, HW40424, HW40423, HW40426, HW40600, HW40465, HW40553, HW40554, HW40383, HW40603, HW40653, HW40655, HW40656, HW40473, HW40555, HW |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0481-2024%22