FDA Device recall Z-0583-2025
Spectral CT and Spectral CT Plus
- FDA Device
- Class II
- ongoing
- Published 2024-12-11
On 2024-12-11 the FDA classified a Class II recall of Spectral CT and Spectral CT Plus by Philips North America, citing potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the pat….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0583-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-11-16 |
| Published | 2024-12-11 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.
Reason
Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID software issue that may lead to misdiagnosis.
Identifiers
| code_info | UDI-DI: (01)00884838101111. Software Version: 5.0.X. Model Number (Serial Numbers): 728333 (10266, 10261, 10264, 10219…UDI-DI: (01)00884838101111. Software Version: 5.0.X. Model Number (Serial Numbers): 728333 (10266, 10261, 10264, 10219, 10265, 10197, 10182, 10175, 10192, 10179, 10196, 10180, 10188, 10190, 10181, 10183, 10178, 10186, 10185, 10199, 10184, 10198, 10193, 10187, 10189, 10195, 10191, 10194, 10018, 10023, 10027, 10062, 10072, 10078, 10079, 10080, 10084, 10088, 10101, 10104, 10112, 10117, 10113, 10116, 10032, 10049, 10133, 10140, 10152, 10153, 10166, 10170, 10171, 10173, 10172, 10174, 10016, 10015, 10017, 10019, 10022, 10021, 10020, 10025, 10024, 10033, 10031, 10034, 10035, 10036, 10038, 10039, 10040, 10041, 10043, 10042, 10014, 10044, 10046, 10048, 10045, 10047, 10051, 10050, 10053, 10055, 10052, 10056, 10058, 10057, 10059, 10060, 10061, 10065, 10066, 10068, 10067, 10070, 10071, 10063, 10069, 10074, 10075, 10077, 10076, 10081, 10082, 10087, 10089, 10085, 10086, 10091, 10094, 10093, 10095, 10096, 10090, 10099, 10098, 10102, 10100, 10103, 10097, 10105, 10106, 10107, 10108, 10109, 396001, 396 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Philips North America
- Ingenia 3.0T and 3.0T CX
FDA Device · Class II · 2026-10-07 · Philips North America - MR 7700
FDA Device · Class II · 2026-10-07 · Philips North America - SmartPath to MR 7700
FDA Device · Class II · 2026-10-07 · Philips North America - Upgrade to MR 7700
FDA Device · Class II · 2026-10-07 · Philips North America - Upgrade dStream
FDA Device · Class II · 2026-10-07 · Philips North America
All Philips North America recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0583-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.