RecallRadar

FDA Device recall Z-0527-2014

Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree C…

On 2013-12-25 the FDA classified a Class II recall of Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree C… by Karl Storz Endoscopy America, citing KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction ch….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0527-2014
ClassificationClass II
Statusterminated
Initiated2013-11-13
Published2013-12-25
Last updated2026-09-13 11:53 UTC
CompanyKarl Storz Endoscopy America
DistributionUS

Product

Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.

Reason

KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.

Identifiers

code_infoC1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0527-2014%22