FDA Device recall Z-0526-2014
Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2. "…
- FDA Device
- Class II
- terminated
- Published 2013-12-25
On 2013-12-25 the FDA classified a Class II recall of Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2. "… by Karl Storz Endoscopy America, citing KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction ch….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0526-2014 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2013-11-13 |
| Published | 2013-12-25 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Karl Storz Endoscopy America |
| Distribution | US |
Product
Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2. " Transoral intubation " Transnasal intubation " Evaluation of placement of endotracheal and double lumen endotracheal tubes " Provide visual access to the larynx and tracheobronchial tree.
Reason
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
Identifiers
| code_info | C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Karl Storz Endoscopy America
- Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree C…
FDA Device · Class II · 2013-12-25 · Karl Storz Endoscopy America - Axillent Debakey Grasping Forceps, Part No. 33821DYD In general the surgeons use 33821DYD Debakey forceps to grasp tissue
FDA Device · Class II · 2013-12-18 · Karl Storz Endoscopy America - 27023WU Balloon Catheter, 3 Fr., 2/pkg, sterile, contains latex
FDA Device · Class II · 2013-07-24 · Karl Storz Endoscopy America - Knife, Cold, Straight (27068K/6)
FDA Device · Class II · 2013-05-22 · Karl Storz Endoscopy America - Electrode, Roller Ball, 24FR (27050NK/6)
FDA Device · Class II · 2013-05-22 · Karl Storz Endoscopy America
All Karl Storz Endoscopy America recalls